Brilliance CT 16 recall
Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core…
Why was Brilliance CT 16 recalled?
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Brilliance CT 16: Model Number: 728246; Serial Number: 5314, 6719, 6726, 6727, 6728, 6729, 6730, 6731, 6732, 6734, 6735, 6736, 6737, 6738, 6744, 6746, 6747, 6748, 6749, 6750, 6751, 6752, 6753, 6754, 6756, 6757, 6758, 6759, 6760, 6761, 6762, 6764, 6767, 6768, 6772, and 50213. Brilliance CT 64: Model Number: 728231; Serial Number: 10805, 10806, 90047, 95880, 95881, 95882, 95885, 95892, 95903, 95907, 95910, 95912, 95913, 95914, 95915, 95918, 95920, 95921, 95922, and 98895. Brilliance CT Big Bore Oncology: Model Number: 728243; Serial Number: 7971, 7977, 7980, 7989, and 7994. Brilliance CT Big Bore Radiology: Model Number: 728244; Serial Number: 7957, 7965, 7979, and 7985. Brilliance iCT: Model Number: 728306; Serial Number: 100477, 100483, 100487, 100489, 100491, 100492, 100494, 100495, 100496, 100500, 100502, 100504, and 100507. Ingenuity CT: Model Number: 728326; Serial Number: 300165 and 333013. Ingenuity Core: Model Number: 728321; Serial Number: 310138, 310142, 310144, 310145, 310149, 310151, 310152, 310153, 310154, 310155, 310158, 310160, 310161, 310162, 310163, 310164, 333013, 333015, and 333017. Ingenuity Core128: Model Number: 728323; Serial Number: 320113, 320114, 320116, 320117, 320118, 320120, 320121, 320122, 320124, 320125, 320128,320129, 320130, 320134, 320137, 320138, 320139, 320140, 320141, 320142, 320143, 320144,320145, 320146, 320147, 320148, 336009, 336010, 336011, 336012, 336013, 336015, 336016, 336018, and 336023. Ingenuity Flex: Model Number: 728317; Serial Number: 345005
- Product
- Brilliance CT 16, Brilliance CT 64, Brilliance CT Big Bore Oncology, Brilliance CT Big Bore Radiology, Brilliance iCT, Ingenuity CT, Ingenuity Core…
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Quantity
- 135
- Distribution
- Nationwide Distribution - USA including the states of CA, FL, IL, IN, KY, MD,SC, and TX.
- Recall date
- April 23, 2014
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)