Monoject Hypodermic Safety Needle recall: Sterility Concern
Monoject Hypodermic Safety Needle 30G X 1/2" Item Code 1183005 - Product Usage: used for general purpose injection and aspiration of fluids from…
Why was Monoject Hypodermic Safety Needle recalled?
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot No. 15063001, Item Code 1183005, Expires 05/31/2020
- Product
- Monoject Hypodermic Safety Needle 30G X 1/2" Item Code 1183005 - Product Usage: used for general purpose injection and aspiration of fluids from…
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 8000
- Distribution
- Worldwide distributions - US Nationwide distributions in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and countries of Canada, Australia, South Africa, Belgium, Germany, Ireland, Portugal Spain, and UK.
- Recall date
- March 18, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)