FDAClass IIJuly 27, 2022

AESCULAP S4 Cervical Cross Connector L 33-42mm recall

AESCULAP S4 Cervical Cross Connector L 33-42mm, S4C CROSS CONNECTOR VARIABLE 33-42MM, product code SW116T. Used in spinal fusion surgery.

Why was AESCULAP S4 Cervical Cross Connector L 33-42mm recalled?

Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: 04038653362332

Product
AESCULAP S4 Cervical Cross Connector L 33-42mm, S4C CROSS CONNECTOR VARIABLE 33-42MM, product code SW116T. Used in spinal fusion surgery.
Recalling firm
Aesculap Implant Systems LLC
Quantity
20
Distribution
United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY.
Recall date
July 27, 2022
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls