LOW Profile Non-lock Screw recall
LOW PROFILE NON-LOCK SCREW; 2.7mm x 22 mm; REF 131227222; LOT RM105E. Provides the orthopaedic surgeon a means of bone fixation and helps generally in…
Why was LOW Profile Non-lock Screw recalled?
A Biomet investigation found that the Low Profile Non-Locking Screw 2.7mmx22mm screw (Part Number: 131227222 Lot: RM105E) was anodized green and should have been anodized gold per the print.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
REF 131227222 LOT RM105E
- Product
- LOW PROFILE NON-LOCK SCREW; 2.7mm x 22 mm; REF 131227222; LOT RM105E. Provides the orthopaedic surgeon a means of bone fixation and helps generally in…
- Recalling firm
- Biomet, Inc.
- Quantity
- 45
- Distribution
- Nationwide Distribution-including the states of CA, IA, NY, FL, PA, UT, NC, SC, KY, CO, MA, GA, WA, TX, OK, and IL.
- Recall date
- April 22, 2015
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)