FDAClass IIApril 25, 2018

T10-NT recall

T10-NT, Model 78107 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or…

Why was T10-NT recalled?

Potential risk for helium gas inside the MR examination room during a magnet quench

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

(Added 6/25/18): 4170 4113 4082 4176 4147 4319 4397 4285 4308 4408 4241 4064 4328 4312 4294 4296 4009 4274 4258 4297 4246 4200 4041 4207 4327 4318 4089 4102 4222 4025 4321 4177 4015 4088 4106 4260 4183 4272 4416 4365 4047 4411 4254 4301 4261 4307 4267 4386 4033 4099 4251 4339 4374 4162 4322 4320 4069 4006 4273 4418 4131 4194 4201 4152 4146 4095 4391

Product
T10-NT, Model 78107 Product Usage: Indicated for use as a diagnostic device. It can produce cross-sectional images, spectroscopic images and/or…
Recalling firm
Philips Electronics North America Corporation
Quantity
8,205 units in total
Distribution
Worldwide Distribution - US Nationwide; International - Albania, Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Chile, China, Colombia, Congo (Democratic Republic of the), Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Republic of Korea - South Korea, Kuwait, Kyrgyzstan, Latvia, Lebanon, Lesotho, Libya, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Mauritius, Mexico, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, Yemen & Zimbabwe.
Recall date
April 25, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls