FDAClass IApril 16, 2014

Draeger Oxylog 3000 plus ventilators recall

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Why was Draeger Oxylog 3000 plus ventilators recalled?

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005

Product
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Recalling firm
Draeger Medical Systems, Inc.
Quantity
3
Distribution
USA Nationwide Distribution in the states of WI and NY.
Recall date
April 16, 2014
Classification
Class I
Status
Terminated
Agency
FDA (device)