Zeiss brand OPMI Pentero C recall
Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and…
Why was Zeiss brand OPMI Pentero C recalled?
Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813
- Product
- Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and…
- Recalling firm
- Carl Zeiss Meditec, Inc.
- Quantity
- 6 units
- Distribution
- US Nationwide Distribution.
- Recall date
- June 12, 2013
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)