LKT200 Perfusion Circuit recall
LKT200 Perfusion Circuit, part of Lifeport Kidney Perfusion Transporter KTR - Product Usage: The circuit is an accessory to LifePort Kidney Perfusion…
Why was LKT200 Perfusion Circuit recalled?
Observed leakage at the sample port assembly which could potentially lead to loss of perfusion solution needed to maintain continuous machine perfusion.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
LKT2000 Lots 119944, 119765, 119648, 120055, 120242.
- Product
- LKT200 Perfusion Circuit, part of Lifeport Kidney Perfusion Transporter KTR - Product Usage: The circuit is an accessory to LifePort Kidney Perfusion…
- Recalling firm
- Organ Recovery Systems, Inc.
- Quantity
- 2274
- Distribution
- Distributed nationwide (32 states) and to Canada.
- Recall date
- March 11, 2020
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)