MEVATRON M 7400/7440(MEX) Accelerator recall
MEVATRON M 7400/7440(MEX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to…
Why was MEVATRON M 7400/7440(MEX) Accelerator recalled?
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Device Model Number 8319758 Serial numbers: 3009 2958 2863 2972 2646 3079 2516 2962 3055 1452 2702 2125 3066 2689 2228 2564 2507 1904 3075 2375 3013 3068 2596 3025 3049 2965 2045 3084 2855 2906 2292 2384 2067 2818 3089
- Product
- MEVATRON M 7400/7440(MEX) Accelerator, Linear, Medical The intended use of the SIEMENS branded PRIMUS family of linear accelerator systems is to…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 35 systems
- Distribution
- Nationwide Distribution
- Recall date
- March 15, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)