Seca 402/403 Baby Scale Cart recall
Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009
Why was Seca 402/403 Baby Scale Cart recalled?
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
serial numbers: 6002142000002 - 6002142001101
- Product
- Seca 402/403 Baby Scale Cart, Ref # 4020000009/ REF # 4030000009
- Recalling firm
- Seca
- Quantity
- 254 scale carts
- Distribution
- US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada
- Recall date
- March 11, 2020
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)