Remel Haemophilus Test Medium(100mm) 10/PK recall
Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative…
Why was Remel Haemophilus Test Medium(100mm) 10/PK recalled?
When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Numbers: Lot 220711 Exp. 2021-2-17
- Product
- Remel Haemophilus Test Medium(100mm) 10/PK, REF: R01503 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative…
- Recalling firm
- Remel Inc
- Quantity
- 19 units
- Distribution
- US Nationwide distribution.
- Recall date
- April 21, 2021
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)