FDAClass IIApril 21, 2021

Remel Haemophilus Test Medium recall

Remel Haemophilus Test Medium (150 mm) 10/PK, REF 04033 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative…

Why was Remel Haemophilus Test Medium recalled?

When tested with H. influenza ATCC 49247 and ATCC 49766 both organisms may exhibit lighter than normal growth making zones difficult to measure.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: 201334 Exp. 2021-01-21, 210892 Exp. 2021-02-03, 217048 Exp. 2021-02-10

Product
Remel Haemophilus Test Medium (150 mm) 10/PK, REF 04033 Remel Haemophilus Test Medium (HTM) is a solid medium recommended for use in qualitative…
Recalling firm
Remel Inc
Quantity
408 units
Distribution
US Nationwide distribution.
Recall date
April 21, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls