FDAClass IIApril 22, 2015

NobRpl Tapered Groovy 6.0mm 6x11.5mm recall

NobRpl Tapered Groovy 6.0mm 6x11.5mm, Article No. 36103. Nobel Biocare's Dental Implants are root-form endosseous implants intended to be surgically…

Why was NobRpl Tapered Groovy 6.0mm 6x11.5mm recalled?

A material used in the packaging of the dental implants may impart a sticky substance on the top surfaces of the implants and in some cases on the first threads.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Batch No. 113010001.

Product
NobRpl Tapered Groovy 6.0mm 6x11.5mm, Article No. 36103. Nobel Biocare's Dental Implants are root-form endosseous implants intended to be surgically…
Recalling firm
Nobel Biocare Usa Llc
Quantity
72 units
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Austria, Australia, Belgium, Bulgaria, Canada, Switzerland, Czech Republic, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Lithuania, Luxembourg, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Sweden, Turkey, and United Arab Emirates.
Recall date
April 22, 2015
Classification
Class II
Status
Terminated