FDAClass IIApril 25, 2018

Milford Smartbase recall

Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Why was Milford Smartbase recalled?

Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Lots

Product
Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B
Recalling firm
Autochair
Distribution
PA, and worldwide
Recall date
April 25, 2018
Classification
Class II
Status
Ongoing
Agency
FDA (device)