FDAClass IIApril 25, 2018

IQon Spectral CT recall

IQon Spectral CT - Model no. 728332, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce…

Why was IQon Spectral CT recalled?

Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial no: 860023 860035 860047 860059 860010 860014 860015 860013 860029 860006 860002 860061 860011 860020 860024 860008 860039 860043 860018 860012 860022 860032 860033 860040 860009 860016 860017 860042 860003 860005 860037 860041 860025 860038 860036 860052 860056 860021 860053 860019 860054 860044 860028 860030 860026 860031 860034

Product
IQon Spectral CT - Model no. 728332, Computed Tomography X-Ray system Product Usage: Computed Tomography X-ray systems intended to produce…
Recalling firm
Philips Medical Systems (Cleveland) Inc
Quantity
47
Distribution
Worldwide Distribution - U.S. Nationwide. International distribution to Canada, Mexico, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Chile, China, Colombia, Croatia, Cuba, Czech Republic, Denmark, Djibouti, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Laos, Lebanon, Lithuania, Macedonia, Malaysia, Maldives, Mali, Martinique, Monaco, Mongolia, Myanmar, Nepal, Netherlands, New Calcedonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Rwanda, Saint Pierre and Miquelon, Samoa, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, and Vietnam.
Recall date
April 25, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)