FDAClass IIFebruary 25, 2026

Medline Convenience Kits: 1) TOTE ABRAMS KNEE recall: Sterility Concern

Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE…

Why was Medline Convenience Kits: 1) TOTE ABRAMS KNEE recalled?

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

1) DYKM1801B, UDI-DI: 10653160360421(each), 00653160360424(case), Lot Number: 2024100790; 2) DYKM1776, UDI-DI: 10888277733794(each), 20888277733791(case), Lot Number: 2025013190; 3) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2024071790; 4) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2024072390; 5) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024080190; 6) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024101490; 7) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2025010890; 8) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2025011090; 9) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2025012890; 10) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2025020690; 11) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2024012590; 12) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024030690; 13) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2024041790; 14) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024052290; 15) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2024053190; 16) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024060590; 17) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2024061890; 18) DYKM1783A, UDI-DI: 10193489111330(each), 20193489111337(case), Lot Number: 2024062890; 19) DYKM1807, UDI-DI: 10193489191219(each), 20193489191216(case), Lot Number: 2024071990; 20) DYKM1781A, UDI-DI: 10653160352730(each), 00653160352733(case), Lot Number: 2024092390; 21) DYKM1796A, UDI-DI: 10653160354888(each), 00653160354881(case), Lot Number: 2024093090; 22) DYKM1977B, UDI-DI: 10653160360797(each), 00653160360790(case), Lot Number: 2024091290; 23) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2024012490; 24) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024022190; 25) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023120590; 26) DYKM1751A, UDI-DI: 10193489111002(each), 20193489111009(case), Lot Number: 2023121190; 27) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023121490; 28) DYKM1776, UDI-DI: 10888277733794(each), 20888277733791(case), Lot Number: 2024031590; 29) DYKM1864, UDI-DI: 10193489191257(each), 20193489191254(case), Lot Number: 2024010880; 30) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023070590; 31) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023071290; 32) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023111590; 33) DYKM1977B, UDI-DI: 10653160360797(each), 00653160360790(case), Lot Number: 2023120890; 34) DYKM1801B, UDI-DI: 10653160360421(each), 00653160360424(case), Lot Number: 2024011790; 35) DYKM1801B, UDI-DI: 10653160360421(each), 00653160360424(case), Lot Number: 2024062890; 36) DYKM1925, UDI-DI: 10193489190311(each), 20193489190318(case), Lot Number: 2024040180; 37) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023072590; 38) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023080190; 39) DYKM1927A, UDI-DI: 10193489199901(each), 20193489199908(case), Lot Number: 2023081490; 40) DYKM1767B, UDI-DI: 10193489112368(each), 20193489112365(case), Lot Number: 2023102790; 41) DYKM1864, UDI-DI: 10193489191257(each), 20193489191254(case), Lot Number: 2023101680; 42) DYKM1977B, UDI-DI: 10653160360797(each), 00653160360790(case), Lot Number: 2023063090; 43) DYKM1925, UDI-DI: 10193489190311(each), 20193489190318(case), Lot Number: 2023112780; 44) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023050490; 45) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023060290; 46) DYKM1807, UDI-DI: 10193489191219(each), 20193489191216(case), Lot Number: 2024032990; 47) CVI4500A, UDI-DI: 10653160327646(each), 00653160327649(case), Lot Number: 2020121550

Product
Medline Convenience Kits: 1) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 2) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 3) TOTE…
Recalling firm
Medline Industries, LP
Quantity
1496 units
Distribution
Worldwide distribution - US Nationwide.
Recall date
February 25, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)