GE Healthcare: a) Prodigy recall
GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734…
Why was GE Healthcare: a) Prodigy recalled?
Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734, LU42344, LU42365 b) Prodigy Advanced, Model Numbers: LU42361, LU42397
- Product
- GE Healthcare: a) Prodigy, Model Numbers: LU7248, LU8905, LU40427, LU40431, LU40626, LU40637, LU40626, LU40637, LU42021, LU42025, LU41730, LU41734…
- Recalling firm
- GE Medical Systems Ultrasound & Primary Care Diagnostics, LL
- Quantity
- 303 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco, Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
- Recall date
- April 25, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)