FDAClass IIMay 29, 2019

AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 recall: Impurity Out of Spec

AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic…

Why was AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 recalled?

Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Number: 21302771, 21302815, 21302786, 21302800, 21302843, 21302928, 21302842, 21302915, 21302900, 21302933, 21302783, 21302922, 21302934, 21302935, 21302942, 21302956, 21302957, 21302989, 21303002, 21303000, 21303001, 21303067, 21303004, 21303044, 21303042, 21302899, 21303091

Product
AURIGA XL 4007 GENERAL SYSTEM UPN: M0068FS4007G0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic…
Recalling firm
Boston Scientific Corporation
Quantity
27 (OUS)
Distribution
Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
Recall date
May 29, 2019
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls