FDAClass IIApril 21, 2021

Covidien Signia Small Diameter Curved Tip Intelligent recall

Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT

Why was Covidien Signia Small Diameter Curved Tip Intelligent recalled?

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: N0A1038UY N0E0016UY N0F0616UY N0K0934UY N0B0493UY N0F0254UY N0F0817UY N0M0334UY N0C0436UY N0F0518UY N0G0388UY N9K1095UY N0D0731UY N0F0615UY N0J0859UY N9L0797UY UDI: 10884521741843 20884521741840

Product
Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT
Recalling firm
Covidien, LP
Quantity
5928 units WW: 4212 US 1716 OUS
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Recall date
April 21, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)