FDAClass IIApril 20, 2016

SYNTHECEL Dura Repair recall: Labeling Error

SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura…

Why was SYNTHECEL Dura Repair recalled?

It was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures post implantation, which may expose patients to potential risks in the event a surgeon intends to temporarily apply and subsequently remove a SYNTHECEL implant. SYNTHECEL Dura Repair labeling and promotional material are being updated to clarify that SYNTHECEL is intended to remain in place and is not intended to be removed.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

SYNTHECEL Dura Repair 2.5 cm x 2.5 cm (1x1), part numbers: SC.400.006.01S, Lot Numbers: 7840311, 7855336, 7860563, 7909569, 7922340, 7932410, 7984104

Product
SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura…
Recalling firm
Synthes (USA) Products LLC
Quantity
212 units
Distribution
Nationwide: NY, PA, AL, WI, FL, MO, TN, OH, MI, MN, CO, CA, NV, AZ, LA, OR, TX, IA, IN, MS, IL, AR, WV, HI, VA, MD, WA, OK, MA, KY, NJ, CT, No foreign accounts.
Recall date
April 20, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls