BD Pyxis: Anesthesia Station ES recall: Software Issue
BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC…
Why was BD Pyxis: Anesthesia Station ES recalled?
Sync timeout may occur between the server and automated dispensing device, due to a server software error which may occur if syncing a large amount of data that takes excessive time, which may cause the dispensing devices to become disconnected from the ES server; dispensing devices will not receive updated information, which may lead to incorrect therapy, and delay in accessing medication.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
BD Pyxis Enterprise Server software versions: 1.7.4.128, 1.7.4.134, 1.7.4.136, 1.7.4.137, All 1.7.1, All 1.7.2, All 1.7.3. REF/UDI-DI: 327/10885403477836, 323/10885403512667, 352/10885403512674, 1116-00/10885403512605
- Product
- BD Pyxis: Anesthesia Station ES, REF: 327, MedStation ES, REF: 323, MedStation ES Tower, REF: 352, CII Safe ES, REF: 1116-00, CII Safe ES Desktop PC…
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 727
- Distribution
- Worldwide - US Nationwide distribution including in the states of KY, OH, NY, MO, NC, PA, FL, TN, CT, LA, MN, VA, MD, CA, TX, NE, IA, OK, MI, PR, SD, IN, ME, KS, IL, NM, WA, AR, HI, MS, GA, WY, WI, VT, AL, NV, NJ, DC, AZ, MT, AK, OR, CO, NH, MA, SC, ID, UT, ND and the countries of TW, CA, AE, MX, SG, AU, NZ, GB, CH, BH.
- Recall date
- March 26, 2025
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)