SOLOPATH Balloon Expandable TransFemoral System recall
SOLOPATH Balloon Expandable TransFemoral System, Model Numbers STFI-1425, STFI-1435, STFI-1625, STFI-1635, STFI-1825, STFI-1835, STFI-1925, STFI-1935…
Why was SOLOPATH Balloon Expandable TransFemoral System recalled?
There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting in a loss of the smooth transition from the surface of the tip to the outer surface of the expandable sheath.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Lot Codes: UG22 UG28A UK02 VA20 VN17 UG22 UK26 VA25 VC15 VG22 VN30 UH29 VA05 VG07 VG26 UH29 VH21 VK29 UH29 UK30 UP20 VF31 WG13 WK30 UH28 UK26 UL28 VD09 VH31 VM19 WP12 UG22 UG22A UH27 UH27A UL14 VA19 VG22 VM24 VP13 WG13 WL28 UG22 UH29 UL14 UL28 UN16 UP20 VA13 VC15 VD09 VH28 VK21 VL06 VL28 VM24 VN21 WG13 XA30 UG22 UL14 UN28 VH27 WM31 UN28 VG02 VP07 WN07
- Product
- SOLOPATH Balloon Expandable TransFemoral System, Model Numbers STFI-1425, STFI-1435, STFI-1625, STFI-1635, STFI-1825, STFI-1835, STFI-1925, STFI-1935…
- Recalling firm
- Terumo Medical Corporation
- Quantity
- 2228
- Distribution
- The products were distributed US nationwide. The products were distributed to the following foreign countries: Australia, Belgium, Brazil, Canada, Portugal, Hong Kong, Peru, Singapore.
- Recall date
- June 5, 2019
- Classification
- Class I
- Status
- Terminated
- Agency
- FDA (device)