FDAClass IIApril 8, 2015

ACCU-CHEK Connect Diabetes Management App recall

ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.

Why was ACCU-CHEK Connect Diabetes Management App recalled?

Roche Diabetes Care has become aware of an issue with the Accu-Chek Connect diabetes management app that could potentially lead to inaccurate bolus advice being provided to the user. A thorough investigation of the situation revealed that this issue may occur if the user changes the screen orientation of the phone from portrait to landscape or vise versa while looking at the Bolus Advisor or Carb

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

N/A

Product
ACCU-CHEK Connect Diabetes Management App; Instruction Manual Designed to transfer data for diabetes management.
Recalling firm
Roche Diabetes Care, Inc.
Quantity
The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Distribution
There are no US distributors as the affected device is not marketed in the US. The affected device was distributed in Germany, Italy and South Africa. The application has been downloaded total of 644 times (Italy-219, South Africa-24, Germany-401) and there are 113 bolus advice activations.
Recall date
April 8, 2015
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls