AquaBplus Water Purification System recall
AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US)…
Why was AquaBplus Water Purification System recalled?
AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502
- Product
- AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US)…
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 2 units
- Distribution
- US Nationwide distribution in the states of NC and WV.
- Recall date
- July 20, 2022
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)