Normed Charcot Osteotome Chisel recall
Normed Charcot Osteotome Chisel (Various sizes)
Why was Normed Charcot Osteotome Chisel recalled?
Product may contain a gap between the handle and metal chisel which can lead to failure to properly disinfect or sterilize between use.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Models: 26.08.506 OSTEO 8MM AND W/HDL 20CM/8IN 26.08.507 OSTEO 10MM AND W/HDL 20CM/8IN 26.08.508 OSTEO 12MM AND W/HDL 20CM/8IN 26.08.509 OSTEO 13MM CRVD R 10 23CM/9IN 26.08.619 OSTEO 19MM CRVD R 10 23CM/9IN (No US distribution) 26.08.712 OSTEO 12MM CRVD R 6 23CM/9IN (No US distribution) 26.08.715 OSTEO 15MM CRVD R 6 23CM/9IN (No US distribution) 26.08.718 CHARC OSTE 18MM CRD R6 23CM/9 26.08.812 OSTEO 12MM CRVD R 8 23CM/9IN (No US distribution) 26.08.815 OSTEO 15MM CRVD R 8 23CM/9IN (No US distribution) 26.08.818 CHARC OSTE 18MM CRD R8 23CM/9 (No US distribution) 26.08.900 ARTHROPIC ARTH 18CM 45TUFF HDL 26.08.910 RUTREK CHIS 20CM/8 DBL CUT BLD
- Product
- Normed Charcot Osteotome Chisel (Various sizes)
- Recalling firm
- Zimmer GmbH
- Distribution
- CO, IN, TX, IL
- Recall date
- April 18, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)