FDAClass IIApril 18, 2018

Bard(R) Premature Infant Feeding Tube recall

Bard(R) Premature Infant Feeding Tube, ENFit(TM), REF EN0036400 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are intended to…

Why was Bard(R) Premature Infant Feeding Tube recalled?

BMD has identified that the product code/lot number combination identified above may have ineffectively fused tubing and ENFit(TM) connector.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: NGAXX216, NGAVX224, NGBN2861

Product
Bard(R) Premature Infant Feeding Tube, ENFit(TM), REF EN0036400 Product Usage: BARD(R) Nasogastric Sump Tubes with ENFit(TM) connector are intended to…
Recalling firm
C.R. Bard, Inc.
Quantity
42200 units
Distribution
US Nationwide Distribution
Recall date
April 18, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls