Abbott ARCHITECT Estradiol Reagent Kit - Product Usage recall

Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the…

Why was Abbott ARCHITECT Estradiol Reagent Kit - Product Usage recalled?

Patient results may be falsely elevated. This patient impact only applies to patients currently being treated with or recently treated with the drug Mifepristone.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

List Number 7K72-20 - ALL LOTS, List Number 7K72-25 - ALL LOTS, and List number 7K72-35 - ALL LOTS.

Product
Abbott ARCHITECT Estradiol Reagent Kit - Product Usage: The ARCHITECT Estradiol assay is a Chemiluminescent Microparticle Immunoassay (CMIA) for the…
Recalling firm
Abbott Ireland Diagnostics Division
Quantity
2,724
Distribution
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR. AZ, CA, CT, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, KS, MT, NC, ND, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, AND PUERTO RICO, and countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Belarus, Belgium, Bosnia and Herzegovi, Brazil, Brunei, Botswana, Canada, Chile, China, Columbia, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Gaza and Jericho, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland,Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Mauritius, Mexico, Moldova, Montenegro, Morocco, Mozambique, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland,Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turks, UAE, Uganda, United Kingdom, and Vietnam.
Recall date
May 22, 2019
Classification
Class II
Status
Terminated