FDAClass IIMay 22, 2019

Molift 2-Point sling bar recall

Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021

Why was Molift 2-Point sling bar recalled?

The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All units manufactured between May 2013 and October 2018. Date of manufacture is labeled on the product.

Product
Molift 2-Point sling bar, an accessory for Molift Mover 205. Item nos. 1430016 1430017 1430021
Recalling firm
Moller Vital
Quantity
427 total
Distribution
Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
Recall date
May 22, 2019
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls