FDAClass IIFebruary 18, 2026

Azurion 5 M12 System recall

Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X

Why was Azurion 5 M12 System recalled?

Under certain conditions, the table may move unexpectedly when the Reset Geometry button is pressed - even when a table lock is active.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

System Model Numbers: (1)722227, (2)722231; UDI-DIs: (1)884838099227, (2)884838099258, (3)884838116740; Serial Numbers: All;

Product
Azurion 5 M12 System Model Numbers: (1)722227, (2)722231; Software Version Number - All versions: R1.X, R2.X, R3.X
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Quantity
326 units (10 US, 316 OUS)
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, FrenchPolynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
Recall date
February 18, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)