FDAClass IIMarch 4, 2020

Ulrich medical neon3 OCT spinal stabilization recall

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Why was Ulrich medical neon3 OCT spinal stabilization recalled?

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot Numbers: U010577, U016620, U019084

Product
Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
Recalling firm
ulrich medical USA Inc
Quantity
179 units
Distribution
AZ, TX, GA, FL, AR, CA
Recall date
March 4, 2020
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls