FDAClass IIApril 13, 2016

VITROS 5600 Integrated System recall

VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety…

Why was VITROS 5600 Integrated System recalled?

The two sets of wires (4 total wires) exiting from the EMI filters were switched, creating a situation where a module that is intended to be de-energized will remain energized only if the associated AC power cord is unplugged.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Codes: 6802413, 6802915 (Refurbished) Serial numbers: J56000024 - J56002387

Product
VITROS 5600 Integrated System, Unique Device Identifier No. 10758750002740 & 10758750007110. Intended for use in the in vitro measurement of a variety…
Recalling firm
Ortho-Clinical Diagnostics
Quantity
2,101 Total units (1,043 domestically & 1,058 internationally)
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Recall date
April 13, 2016
Classification
Class II
Status
Terminated
Agency
FDA (device)