Ysio X.pree recall
Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body including the skull, chest…
Why was Ysio X.pree recalled?
A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Serial Numbers: 10041 10046 10055 10057 10053 10049 10043 10039 10054 10048 10042 10045
- Product
- Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body including the skull, chest…
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 12 units
- Distribution
- US Nationwide distribution in the states of FL, MN, OH, PA, TX.
- Recall date
- April 14, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)