Medtronic HVAD Monitor Data Cable recall
Medtronic HVAD Monitor Data Cable, Product Number: 1575
Why was Medtronic HVAD Monitor Data Cable recalled?
There is the potential for the Controller power cables (AC Adapter, DC Adapter, and Battery cables), HVAD Alarm Adapter, and HVAD Monitor data cables to cause damage to the HVAD Controller ports when inserting the cables into the Controller ports, due to wear from misalignment of the cables to the port.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
All lot codes
- Product
- Medtronic HVAD Monitor Data Cable, Product Number: 1575
- Recalling firm
- Heartware, Inc.
- Quantity
- 1337 units (1110 US)
- Distribution
- Worldwide
- Recall date
- April 14, 2021
- Classification
- Class I
- Status
- Ongoing
- Agency
- FDA (device)