MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG recall
MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG -Used for the percutaneous introduction of a guidewire into the vascular system. Catalog…
Why was MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG recalled?
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI: 15051684022972 UPN: H965457541 Lot Number: Lot: 5784134 5787807 5787808 5787809 5787810 5789980 5790676 5800402 5803531
- Product
- MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG -Used for the percutaneous introduction of a guidewire into the vascular system. Catalog…
- Recalling firm
- Angiodynamics, Inc.
- Quantity
- 1676 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
- Recall date
- March 27, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)