MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" recall
MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog…
Why was MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" recalled?
Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
UDI: 15051684022941 UPN: H965457511 Lot Number: 5789981 5789983 5791592 5791595 5795873 5797462
- Product
- MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog…
- Recalling firm
- Angiodynamics, Inc.
- Quantity
- 959 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
- Recall date
- March 27, 2024
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)