BIOCARE Medical recall: Software Issue
BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1
Why was BIOCARE Medical recalled?
Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Affected US Serial Numbers: BCIP01-0021, BCIP01-0033, BCIP01-0090, BCIP01-0106, BCIP01-0112, BCIP01-0148, BCIP01-0151, BCIP01-0187, BCIP01-0189, BCIP01-0198, BCIP01-0201, BCIP01-0216, BCIP01-0221, BCIP01-0249, BCIP01-0262, BCIP01-0265, BCIP01-0266, BCIP01-0268, BCIP01-0279, BCIP01-0285, BCIP01-0307, BCIP01-0314, BCIP01-0319, BCIP01-0328, BCIP01-0341, BCIP01-0346, BCIP01-0354, BCIP01-0386, BCIP01-0387, BCIP01-0388, BCIP01-0390, BCIP01-0395, BCIP01-0396, BCIP01-0413, BCIP01-0426, BCIP01-0430, BCIP01-0431, BCIP01-0432, BCIP01-0434, BCIP01-0443, BCIP01-0450, BCIP01-0463, BCIP01-0464, BCIP01-0467, BCIP01-0482, BCIP01-0484, BCIP01-0487, BCIP01-0491, BCIP01-0492, BCIP01-0495, BCIP01-0500, BCIP01-0502, BCIP01-0508, BCIP01-0517, BCIP01-0526, BCIP01-0534, BCIP01-0543, BCIP01-0545, BCIP01-0546, BCIP01-0550, BCIP01-0553, BCIP01-0555, BCIP01-0560, BCIP01-0562, BCIP01-0570, BCIP01-0571, BCIP01-0572, BCIP01-0573 Affected OUS serial numbers: BCIP01-0454, BCIP01-0515, BCIP01-0538, BCIP01-0556
- Product
- BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1
- Recalling firm
- Biocare Medical, LLC
- Quantity
- 72 units
- Distribution
- US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands
- Recall date
- April 7, 2021
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)