FDAClass IIApril 7, 2021

Philips SureSigns VM4 recall

Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of…

Why was Philips SureSigns VM4 recalled?

Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Software version: SW A.02.63

Product
Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple physiological parameters of…
Recalling firm
Philips North America, LLC
Quantity
31773 units
Distribution
International distribution in the country of China.
Recall date
April 7, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls