FDAClass IIIFebruary 26, 2020

TLC-2000 Therapeutic Medical Laser System - Product recall

TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.

Why was TLC-2000 Therapeutic Medical Laser System - Product recalled?

Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 3275 to 3350 (inclusive)

Product
TLC-2000 Therapeutic Medical Laser System - Product Usage: Indicated for the temporary relief of knee pain.
Recalling firm
Theralase Inc.
Quantity
76 units
Distribution
Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.
Recall date
February 26, 2020
Classification
Class III
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls