Fresenius 2008 Series Hemodialysis Delivery Systems recall
Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System…
Why was Fresenius 2008 Series Hemodialysis Delivery Systems recalled?
Added Warning Statement: Fresenius 2008 Series Hemodialysis Machines False Blood Leak Alarm When Dialyzing Patients Treated with Hydroxocobalamin (or any form of Vitamin B-12)
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System Fresenius 2008K@home Hemodialysis Machine with bibag System 2008T Hemodialysis Machine
- Product
- Fresenius 2008 Series Hemodialysis Delivery Systems: 2008T, 2008K2, 2008K and 2008K@home: Models: Fresenius 2008K and 2008K2 Dialysate Delivery System…
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Quantity
- 8384 units (8369 US) (15 OUS)
- Distribution
- Nationwide Distribution
- Recall date
- April 13, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)