FDAClass IIApril 18, 2018

Progressa Bed - AC-powered adjustable hospital bed recall

Progressa Bed - AC-powered adjustable hospital bed

Why was Progressa Bed - AC-powered adjustable hospital bed recalled?

A component used within the braking system may break resulting in one or more brake pedals on the bed staying in brake or not being engaged which could result in unwanted bed movement and possible patient or user injury.

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

S159AW9192 through S305AW2321 Beds manufactured from June 8, 2017 to November 1, 2017

Product
Progressa Bed - AC-powered adjustable hospital bed
Recalling firm
Hill-Rom, Inc.
Quantity
2117
Distribution
Worldwide distribution. US nationwide, Canada, Chile, Colombia, Costa Rica, France, Hong Kong, Israel, Jamaica, Japan, Jordan, Malaysia, Mexico, Oman, Panama, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, UAE, and UK.
Recall date
April 18, 2018
Classification
Class II
Status
Terminated
Agency
FDA (device)