FDAClass IIMarch 27, 2024

Assembly recall

Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Why was Assembly recalled?

There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

UDI-DI: M9461091010. Lot Numbers: 2316701, 2232601, 23030, 2228501, 2327802, 22152

Product
Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System
Recalling firm
Think Surgical, Inc.
Quantity
16
Distribution
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.
Recall date
March 27, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)