FDAClass IIApril 7, 2021

EPIX Therapeutics DiamondTemp Unidirectional Large recall: Labeling Error

EPIX Therapeutics DiamondTemp Unidirectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF: CEDT200L…

Why was EPIX Therapeutics DiamondTemp Unidirectional Large recalled?

It has been identified that the catheter size on the packaging labeling does not correspond to the widest part of the catheter. In addition the IFU contains the incorrect diameter of the catheter. This could result in blood stagnation, clots or air emboli.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

All Lots

Product
EPIX Therapeutics DiamondTemp Unidirectional Large Curve Ablation Catheter, 7.5 French Size(2.48 mm), Connector Type: ODU 19 PIN, REF: CEDT200L…
Recalling firm
EPIX THERAPEUTICS, INC
Quantity
60 units
Distribution
Product was not distributed in the US. OUS: Germany, Spain, France, Belgium, United Kingdom, Slovenia, Italy, Canada
Recall date
April 7, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls