FDAClass IIFebruary 18, 2026

Spectral CT recall

Spectral CT; Product Code (REF): 728333;

Why was Spectral CT recalled?

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Product Code (REF): 728333; UDI-DI: 00884838101111; Serial Numbers: 10055, 10075, 10086, 10096, 10106, 10116, 10126, 10136, 10146, 10156, 10166, 10176, 10186, 10196, 10206, 10217, 10227, 10237, 10247, 10257, 10267, 10277, 10289, 10302, 10314, 12009, 12019, 12029, 396013, 10056, 10076, 10087, 10097, 10107, 10117, 10127, 10137, 10147, 10157, 10167, 10177, 10187, 10197, 10207, 10218, 10228, 10238, 10248, 10258, 10268, 10278, 10290, 10303, 12000, 12010, 12020, 12030, 398003, 10057, 10077, 10088, 10098, 10108, 10118, 10128, 10138, 10148, 10158, 10168, 10178, 10188, 10198, 10208, 10219, 10229, 10239, 10249, 10259, 10269, 10279, 10291, 10304, 12001, 12011, 12021, 12031, 398004, 10058, 10078, 10089, 10099, 10109, 10119, 10129, 10139, 10149, 10159, 10169, 10179, 10189, 10199, 10209, 10220, 10230, 10240, 10250, 10260, 10270, 10280, 10292, 10305, 12002, 12012, 12022, 396005, 10059, 10079, 10090, 10100, 10110, 10120, 10130, 10140, 10150, 10160, 10170, 10180, 10190, 10200, 10210, 10221, 10231, 10241, 10251, 10261, 10271, 10281, 10293, 10306, 12003, 12013, 12023, 396006, 10060, 10080, 10091, 10101, 10111, 10121, 10131, 10141, 10151, 10161, 10171, 10181, 10191, 10201, 10211, 10222, 10232, 10242, 10252, 10262, 10272, 10284, 10294, 10307, 12004, 12014, 12024, 396008, 10061, 10081, 10092, 10102, 10112, 10122, 10132, 10142, 10152, 10162, 10172, 10182, 10192, 10202, 10212, 10223, 10233, 10243, 10253, 10263, 10273, 10285, 10298, 10308, 12005, 12015, 12025, 396009, 10062, 10082, 10093, 10103, 10113, 10123, 10133, 10143, 10153, 10163, 10173, 10183, 10193, 10203, 10213, 10224, 10234, 10244, 10254, 10264, 10274, 10286, 10299, 10310, 12006, 12016, 12026, 396010, 10070, 10083, 10094, 10104, 10114, 10124, 10134, 10144, 10154, 10164, 10174, 10184, 10194, 10204, 10214, 10225, 10235, 10245, 10255, 10265, 10275, 10287, 10300, 10311, 12007, 12017, 12027, 396011, 10074, 10085, 10095, 10105, 10115, 10125, 10135, 10145, 10155, 10165, 10175, 10185, 10195, 10205, 10215, 10226, 10236, 10246, 10256, 10266, 10276, 10288, 10301, 10312, 12008, 12018, 12028, 396012,

Product
Spectral CT; Product Code (REF): 728333;
Recalling firm
Philips North America Llc
Quantity
283 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam.
Recall date
February 18, 2026
Classification
Class II
Status
Ongoing
Agency
FDA (device)