Brand Name: Molift Product Name: Molift Mover 300 recall
Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7)…
Why was Brand Name: Molift Product Name: Molift Mover 300 recalled?
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
(1) Model/Catalog Number: M1501; UDI-DI: TBD; Serial Numbers: 601005, 601002, 611506, 621101, 621106, 621104, 621105, 621102, 611505, 611504, 611503, 611502, 611501, 601008, 611507, 601007, 601006, 601001, 621103, 601003, 601004; (2) Model/Catalog Number: M1502; UDI-DI: TBD; Serial Numbers: 611409, 611408, 611407, 611406, 611405, 611404, 611403, 611402, 611401, 611412, 611411, 611410; (3) Model/Catalog Number: M1504; UDI-DI: TBD; Serial Numbers: 600902, 600901, 611303, 611302, 611301; (4) Model/Catalog Number: M15050; UDI-DI: TBD; Serial Numbers: 586003, 586001, 585905, 585903, 585902, 585901, 598605, 586101, 585904, 603602, 591401, 586002, 586006, 586005, 586004, 598608, 598607, 585906, 591601, 598606, 598604, 598602, 598601, 596501, 603601, 598609, 598603; (5) Model/Catalog Number: M15151; UDI-DI: TBD; Serial Numbers: 599703, 599702, 599701; (6) Model/Catalog Number: M15001; UDI-DI: TBD; Serial Numbers: 584401; (7) Model/Catalog Number: M15002; UDI-DI: TBD; Serial Numbers: 591501, 591502;
- Product
- Brand Name: Molift Product Name: Molift Mover 300 Model/Catalog Numbers: (1) M1501, (2) M1502, (3) M1504, (4) M15050, (5) M15151, (6) M15001, (7)…
- Recalling firm
- ETAC A/S
- Quantity
- 71
- Distribution
- Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
- Recall date
- February 11, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)