FDAClass IIMarch 27, 2024

NautilusTM ECMO Oxygenator with Balance Biosurface-used recall

NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and…

Why was NautilusTM ECMO Oxygenator with Balance Biosurface-used recalled?

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

GTIN: (1) 10854916006888; (2) 10854916006901,20854916006908 (New GTIN (USA only)After reboxing). All Serial Numbers between 481022923 and 481024039

Product
NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and…
Recalling firm
Mc3 Inc
Quantity
299 units
Distribution
US Nationwide distribution.
Recall date
March 27, 2024
Classification
Class II
Status
Ongoing
Agency
FDA (device)