FDAClass IIMarch 31, 2021

VITEK 2 Compact recall: Software Issue

VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated…

Why was VITEK 2 Compact recalled?

bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. Each instrument has unique serial number. (System: VITEK 2 Compact 60, Reference Number: 27560, Catalog Number: C57, UDI: 03573026136796); (System: VITEK 2 Compact 30, Reference Number: 27560R, Catalog Number: C57, UDI: 03573026147501); (System: VITEK 2 Compact 30, Reference Number: 27530, Catalog Number: C57, UDI: 03573026136789); (System: VITEK 2 Compact 30, Reference Number: 27530R, Catalog Number: C57, UDI: 03573026147495); (System: VITEK 2 Compact 15, Reference Number: 27415, Catalog Number: C57, UDI: 03573026220624); (System: VITEK 2 Compact 15, Reference Number: 27415R, Catalog Number: C57, UDI: 03573026220631)

Product
VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated…
Recalling firm
Biomerieux Inc
Quantity
6777 systems
Distribution
Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.
Recall date
March 31, 2021
Classification
Class II
Status
Terminated
Agency
FDA (device)
Open FDA enforcement record Latest recalls