FDR Visionary Suite recall
FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
Why was FDR Visionary Suite recalled?
It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be supplied to the power supply circuit even if the power of the device is cut off. As a result, it cannot be denied that the circuit in question overheats and damages the surrounding parts. It may also cause smoke.
What should I do if I have this product?
The recall is ongoing. Follow the firm and FDA instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Model Number: CH-200; (1) Version: 566-16130-23; UDI-DI: 04540217052226; Serial numbers: MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001; (2) Version: 566-16130-31; UDI-DI: 04540217057436; Serial numbers: MP95AA783001; (3) Version: 566-16130-33; UDI-DI: 04540217057450; Serial numbers: MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
- Product
- FDR Visionary Suite; Model Number: CH-200; Version: (1) 566-16130-23, (2) 566-16130-31, (3)
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Quantity
- 16 units
- Distribution
- US Nationwide distribution in the states of AK, CA, CT, FL, IL, NH, OH, WI.
- Recall date
- February 11, 2026
- Classification
- Class II
- Status
- Ongoing
- Agency
- FDA (device)