FDAClass IIMarch 31, 2021

S-Series Equipment Management System recall

S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good…

Why was S-Series Equipment Management System recalled?

The manufacturer has become aware that there is a likelihood that the S-Series Equipment Management System Capacitive Touch Controls on the MFRs rotational electrical magnetic brakes may disengage when an electrocautery device is in use.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Serial Numbers: 2258046 2863490

Product
S-Series Equipment Management System; Model Number(s):0682400550; UDI: 07613327400502 Components that are affected within the S-Series Finished Good…
Recalling firm
Stryker Communications
Quantity
1,393 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV, and WY. The country of Canada.
Recall date
March 31, 2021
Classification
Class II
Status
Ongoing
Agency
FDA (device)