LeadCare II Blood Lead Analyzer recall
LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
Why was LeadCare II Blood Lead Analyzer recalled?
The incorrect placement of AA batteries into certain LeadCare II Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
Part Numbers: 70-6760, 70-6529-R, 70-6760-P, and 70-6760-R. All LeadCare II and LeadCare Plus Blood Lead Analyzers with battery boards versions B and O.
- Product
- LeadCare II Blood Lead Analyzer Part Numbers: 70-6760, 70-6529-R, 70-6760-P, 70-6760-R. Intended for in vitro (external) use only.
- Recalling firm
- Magellan Diagnostics, Inc.
- Quantity
- 6,143
- Distribution
- Worldwide Distribution
- Recall date
- April 11, 2018
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)