FDAClass IIMarch 12, 2025

Tissue Approximation System recall

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

Why was Tissue Approximation System recalled?

Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.

What should I do if I have this product?

The recall is ongoing. Follow the firm and FDA instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Model/UDI-DI/Lot(Expiration): T-4000/00197644217267/230403(2024/04/20); T-5000/00860011435921/240801(2025/08/28), 240901(2025/09/25); T-LAP/00860011435914, 00197644927241/240702(2025/07/08), 230906(2024/11/29), 230905(2024/11/29)

Product
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Recalling firm
TAS Medical Inc
Quantity
114
Distribution
US Nationwide distribution in the states of LA, OK, TN.
Recall date
March 12, 2025
Classification
Class II
Status
Ongoing
Agency
FDA (device)
Open FDA enforcement record Latest recalls